Warning: Electrical Shock Hazard. There are electrically live parts
within the warming unit when it is connected to a power source, even
when the hand-held temperature controller is in the OFF position.
Please forward to
Biomedical Engineering Department
Bair Paws
®
Model 850
Warming Unit
Service
Manual
Vue de la page 0
1 2 3 4 5 6 ... 25 26

Résumé du contenu

Page 1 - Please forward to

Warning: Electrical Shock Hazard. There are electrically live parts within the warming unit when it is connected to a power source, even when the hand

Page 2 - Table of Contents

Bair Paws® Model 850 Warming Unit Service Manual8Mounting the Warming Unit with a Rail-mount to the WallUsing the wall-screw holes on the rail-to-wall

Page 3 - Bair Paws Warming Gowns

Arizant Healthcare Inc.9Using the Temperature Controller Holdere controller holder helps to prevent cord clutter around the patient or accidental dam

Page 4 - Warnings

Bair Paws® Model 850 Warming Unit Service Manual10Service ProceduresCalibrating the Operating TemperaturesService FrequencyCalibrate the unit before p

Page 5 - Safety Inspection

Arizant Healthcare Inc.11MethodTurn the warming unit “ON” and set the controller to its maximum (warmest) setting.1. Press the clip buttons on the hos

Page 6

Bair Paws® Model 850 Warming Unit Service Manual12Extend the hose straight to its full length. Remove the two screws that attach the case bottom 3. (d

Page 7 - Setscrew

Arizant Healthcare Inc.13Replacing the Temperature Controller and/or HoseTools/Equipment NeededMedium Phillips screwdriver•Needle-nose pliers•Method

Page 8

Bair Paws® Model 850 Warming Unit Service Manual14Using either a replacement temperature controller assembly or a replacement hose, thread the 8. cont

Page 9

Arizant Healthcare Inc.15Route the new power cord through the enclosure hole and around the case as shown. Make sure 10. the cord is held firmly by th

Page 10

Bair Paws® Model 850 Warming Unit Service Manual16Cleaning the Warming UnitService FrequencyAs needed.Tools/Equipment NeededSoft cloth•Mild cleaning

Page 11

Arizant Healthcare Inc.17Technical Support and Customer ServiceU.S. Customer ServiceTEL: FAX:800-733-7775 800-775-0002952-947-1200 952-947-1400When

Page 12 - Service Procedures

Table of ContentsIntroduction to the Bair Paws® Patient Adjustable Warming System ...1Important Information about the

Page 13 - Tools/Equipment Needed

Bair Paws® Model 850 Warming Unit Service Manual18Bair Paws Warming Unit Limited WarrantyTwo-Year Limited Warrantya, cArizant Healthcare Inc. (“Compan

Page 14 - Replacing the Filter

Arizant Healthcare Inc.19SpecificationsPhysical Characteristics     11 in. high x 2.5 in. deep x 7.7 in. wide 27.9 cm high x 6

Page 15

Bair Paws® Model 850 Warming Unit Service Manual20Specifications - ContinuedElectrical Characteristics  Airflow: 7-9 cfm  P

Page 16 - Replacing the Power Cord

Arizant Healthcare Inc.21Maintenance Log Date Maintenance Action Performed

Page 17 - Reassembling the Warming Unit

Bair Paws® Model 850 Warming Unit Service Manual22Maintenance Log Date Maintenance Action Performed

Page 18 - Cleaning the Warming Unit

Arizant Healthcare Inc.23Definition of Symbolse following symbols may appear on the product, on the exterior packaging, or in the product labeling. O

Page 19 - U.S. Customer Service

Arizant Healthcare Inc.,10393 West 70th Street,Eden Prairie, MN 55344 USATEL 800-733-7775 • 952-947-1200FAX 800-775-0002 • 952-947-1400www.bairpaws.c

Page 20 - Limitation of Remedies

Arizant Healthcare Inc.1Introduction to the Bair Paws® Patient Adjustable Warming SystemThe Bair Paws patient adjustable warming system consists of t

Page 21 - Specifications

Bair Paws® Model 850 Warming Unit Service Manual2Important Information about the Bair Paws Model 850 Warming Unit Indications/Intended Usee Model 85

Page 22 - Specifications - Continued

Arizant Healthcare Inc.3PrecautionsDo not initiate patient warming unless the Model 850 warming unit is safely placed on a hard •surface or securely

Page 23 - Maintenance Log

Bair Paws® Model 850 Warming Unit Service Manual4Preparing the Bair Paws Model 850 Warming Unit for UseBefore using the warming unit, make sure that i

Page 24

Arizant Healthcare Inc.5Mounting the Warming Unit to the Wall-mountSlide the V-shaped bracket on the back of the warming unit into the groove of the

Page 25 - Definition of Symbols

Bair Paws® Model 850 Warming Unit Service Manual6Mounting the Warming Unit on an IV PoleCa u t i o n : To prevent tipping, mount the Model 850 warming

Page 26

Arizant Healthcare Inc.7Mounting the Warming Unit on a BedrailN: e wall-mount bracket is shipped attached to the rail-mount.Slide the V-shaped bra