Warning: Electrical Shock Hazard. There are electrically live parts within the warming unit when it is connected to a power source, even when the hand
Bair Paws® Model 850 Warming Unit Service Manual8Mounting the Warming Unit with a Rail-mount to the WallUsing the wall-screw holes on the rail-to-wall
Arizant Healthcare Inc.9Using the Temperature Controller Holdere controller holder helps to prevent cord clutter around the patient or accidental dam
Bair Paws® Model 850 Warming Unit Service Manual10Service ProceduresCalibrating the Operating TemperaturesService FrequencyCalibrate the unit before p
Arizant Healthcare Inc.11MethodTurn the warming unit “ON” and set the controller to its maximum (warmest) setting.1. Press the clip buttons on the hos
Bair Paws® Model 850 Warming Unit Service Manual12Extend the hose straight to its full length. Remove the two screws that attach the case bottom 3. (d
Arizant Healthcare Inc.13Replacing the Temperature Controller and/or HoseTools/Equipment NeededMedium Phillips screwdriver•Needle-nose pliers•Method
Bair Paws® Model 850 Warming Unit Service Manual14Using either a replacement temperature controller assembly or a replacement hose, thread the 8. cont
Arizant Healthcare Inc.15Route the new power cord through the enclosure hole and around the case as shown. Make sure 10. the cord is held firmly by th
Bair Paws® Model 850 Warming Unit Service Manual16Cleaning the Warming UnitService FrequencyAs needed.Tools/Equipment NeededSoft cloth•Mild cleaning
Arizant Healthcare Inc.17Technical Support and Customer ServiceU.S. Customer ServiceTEL: FAX:800-733-7775 800-775-0002952-947-1200 952-947-1400When
Table of ContentsIntroduction to the Bair Paws® Patient Adjustable Warming System ...1Important Information about the
Bair Paws® Model 850 Warming Unit Service Manual18Bair Paws Warming Unit Limited WarrantyTwo-Year Limited Warrantya, cArizant Healthcare Inc. (“Compan
Arizant Healthcare Inc.19SpecificationsPhysical Characteristics 11 in. high x 2.5 in. deep x 7.7 in. wide 27.9 cm high x 6
Bair Paws® Model 850 Warming Unit Service Manual20Specifications - ContinuedElectrical Characteristics Airflow: 7-9 cfm P
Arizant Healthcare Inc.21Maintenance Log Date Maintenance Action Performed
Bair Paws® Model 850 Warming Unit Service Manual22Maintenance Log Date Maintenance Action Performed
Arizant Healthcare Inc.23Definition of Symbolse following symbols may appear on the product, on the exterior packaging, or in the product labeling. O
Arizant Healthcare Inc.,10393 West 70th Street,Eden Prairie, MN 55344 USATEL 800-733-7775 • 952-947-1200FAX 800-775-0002 • 952-947-1400www.bairpaws.c
Arizant Healthcare Inc.1Introduction to the Bair Paws® Patient Adjustable Warming SystemThe Bair Paws patient adjustable warming system consists of t
Bair Paws® Model 850 Warming Unit Service Manual2Important Information about the Bair Paws Model 850 Warming Unit Indications/Intended Usee Model 85
Arizant Healthcare Inc.3PrecautionsDo not initiate patient warming unless the Model 850 warming unit is safely placed on a hard •surface or securely
Bair Paws® Model 850 Warming Unit Service Manual4Preparing the Bair Paws Model 850 Warming Unit for UseBefore using the warming unit, make sure that i
Arizant Healthcare Inc.5Mounting the Warming Unit to the Wall-mountSlide the V-shaped bracket on the back of the warming unit into the groove of the
Bair Paws® Model 850 Warming Unit Service Manual6Mounting the Warming Unit on an IV PoleCa u t i o n : To prevent tipping, mount the Model 850 warming
Arizant Healthcare Inc.7Mounting the Warming Unit on a BedrailN: e wall-mount bracket is shipped attached to the rail-mount.Slide the V-shaped bra